A&A 411, L197-L201 (2003)
DOI: 10.1051/0004-6361:20031210
A. J. Bird 1 - A. Bazzano 2 - C. Ferguson 1 - G. La Rosa 3 - G. Malaguti 4 - P. Ubertini 2
1 - School of Physics and Astronomy, University of Southampton,
Highfield, Southampton, SO17 1BJ, UK
2 -
Istituto di Astrofisica Spaziale e Fisica Cosmica, IASF/CNR, Rome, Italy
3 -
Istituto di Astrofisica Spaziale e Fisica Cosmica, IASF/CNR, Sezione di Palermo, Palermo, Italy
4 -
Istituto di Astrofisica Spaziale e Fisica Cosmica, IASF/CNR, Sezione di Bologna, Bologna, Italy
Received 11 July 2003 / Accepted 4 August 2003
Abstract
We describe the in-flight performance of the on-board
calibration unit for the IBIS telescope on INTEGRAL. Both
intrinsic performance and the quality of the calibration
signals provided to the IBIS detector planes are
discussed. The calibration unit intrinsic performance is
assessed based on the diagnostic information in IBIS
housekeeping. The flux of tagged photons, i.e. those arriving
at the detector planes in coincidence with a calibration
strobe, is assessed from in-flight data, and analysed in
conjunction with a detailed simulation. Proposed usage of the
tagged photon flux is discussed, and the expected calibration
accuracy derived. The effect on science data from untagged
calibration unit photons is assessed based on the predicted
rate and distribution of such photons.
Key words: INTEGRAL - gamma-ray - calibration
The IBIS telescope (Ubertini et al. 1996, 2003) on INTEGRAL (Winkler et al. 2003) includes a small in-flight calibration system designed to allow monitoring of the gains and overall performance of the individual detector components making up the ISGRI (Lebrun et al. 2003) and PICsIT (Labanti et al. 2003) detector planes and the active veto system (Quadrini et al. 2003). Here we describe the in-flight performance of the IBIS on-board calibration unit derived during telescope commissioning operations.
The IBIS on-board calibration unit (OBCU) provides a continuous but small flux of 511 and 1275 keV photons to the detector planes, together with a logic signal (hereafter the "calibration strobe'') which is interpreted by the detector planes as a signal that a calibration photon has been emitted. The calibration photons are thus recognised by the detector planes as coming from the OBCU due to their coincidence with the calibration strobe, and are diverted to dedicated calibration data streams, preventing contamination of the science data.
![]() |
Figure 1: Principle of IBIS OBCU operation: photons emitted towards the IBIS detector planes are "tagged'' by simultaneous detections in the BGO tagging system. |
| Open with DEXTER | |
The process of associating a calibration strobe with each emitted
calibration photon is described as "tagging'' the photon. The OBCU
consists of an active tagging system based on a modified IBIS veto
detector and a low intensity (0.4
Ci at launch) 22Na
radioactive source (Poulsen 2001). The 22Na source
disintegrations result in either two 511 keV and one 1275 keV photons
(90% of the time) or just a 1275 keV photon (10% of the time). Since
the two 511 keV photons are emitted 180
apart, the detection
of one photon can be used to indicate a source disintegration while
maintaining the photon flux towards the detector planes (see
Fig. 1).
![]() |
Figure 2: Internal diagnostic pulse height spectra from the OBCU for (left) in-flight operation and (right) during ground calibrations. The individual Gaussian fits to the 511, 1275 and 1786 lines, together with an exponential background component, are shown together with the recorded spectrum. |
| Open with DEXTER | |
The tagging system itself is based on a modified veto detector unit,
which uses a 2 cm thick bismuth germanate (BGO) scintillator block as
the active detection volume, viewed by two photomultipliers. The
standard veto system electronics are modified, and an additional high
voltage unit included, to provide the necessary redundancy against the
failure of one photomultipler or high voltage supply. This detector
system generates a calibration strobe, which is distributed to the
detector planes, whenever the 22Na calibration source generates
an energy deposit between 100-2000 keV in the tagging system. Any
events in the detector planes which are in coincidence with the 2
s calibration strobe are recorded as calibration events.
First activation commenced after 17 days in orbit, in order to allow
for outgassing around the critical high voltage (HV) components. OBCU
health was monitored via instrument housekeeping data. The BGO/PMT
temperature was typically around
C and the HV supply
typically -1100 V.
The count rate in the OBCU tagging detector system, as reported in the instrument housekeeping data was 6120 c/s. This compares to a typical value of 9940 c/s observed during ground calibrations, and so indicated a significant deviation from the expected performance.
Instrument housekeeping also provides for the periodic transmission of
a pulse height spectrum from the OBCU tagging system, which allows
gain monitoring of the active components of the OBCU.
Figure 2(left) shows a typical in-flight spectrum
from the OBCU tagging detector. The three peaks visible in the
spectrum correspond to the expected 511 keV, 1275 keV and 1786 (511+1275) keV peaks from the 22Na source. However, comparing
this to a typical spectrum recorded during ground calibration
(Fig. 2, right), the in-flight spectrum shows
significant departures from the on-ground behaviour,which may be
considered as nominal. Note that the on-ground spectrum is taken in
coincidence with the calibration strobe itself. This superimposes the
thresholds for calibration strobe generation on the spectrum and
causes the sharp discontinuities at pulse height channels
20 and
220.
The most significant difference is the broadening and shift of the 511 keV peak to a lower channel. In the ground spectrum, the lower side of the 511 keV peak is well above the tagging threshold, while in the flight spectrum, the threshold cuts into the peak. Conversely the two high-energy peaks are still in approximately the same channels.
Inspection of the channel-energy relations derived by fitting to the three peaks show that the relationship remains linear, but the peak shift is actually indicative of a large offset being present in the flight configuration. The total number of counts in the flight spectrum indicates a significantly lower BGO count rate compared to the ground data, which is consistent with the reduction seen in the housekeeping counters.
The interpretation of this in-flight behaviour is that the tagging performance is significantly worse than that seen during ground calibrations. The most likely explanation is that the BGO tagging system is being degraded by the presence of a large high-energy particle flux, as seen in the other detectors (Segreto 2003). The modifications made to the veto electronics to avoid "blinding'' was not made on the OBCU due to incompatibility with the need to support cold redundancy in the OBCU.
The high energy flux is creating two linked effects, which are illustrated in Fig. 3, and are sufficient to explain the deviations from nominal behaviour seen in the calibration unit performance reported in the housekeeping data.
![]() |
Figure 3: Proposed saturation, blinding and baseline restoration effects in OBCU electronics due to the detection of high energy photons or charged particles. |
| Open with DEXTER | |
Both effects, of blinding and spectrum shift, have the effect of reducing the tagging efficiency of the calibration unit. The blinding creates a simple dead-time in the tagging system, a fraction of time in which the system cannot generate strobes. The spectrum shift means that low energy events in the tagging system (including some of the important 511 keV events) are shifted below the preset threshold in the tagging system, even when it is set to the minimum allowed value.
The combination of these effects produces an effective dead time in
the tagging detector which can be estimated by comparison of the
ground and in-flight count rates as:
Simulations using the GEANT4 package (Agostinelli 2003) have been used in order to estimate the event rates in the IBIS detector planes as a result of calibration source disintegrations. The calculations include the effects of (a) random coincidences between the detector plane background and OBCU tagging system and (b) the OBCU blinding estimated in the previous section.
The effect of random coincidences is to remove science events from the science data streams; the OBCU creates an effective dead-time for science data and simultaneously degrades the quality of the calibration data, which has a random sample of detector counts added to it.
Conversely, the main effect of blinding and/or loss of tagging efficiency is to divert events from the calibration data-streams into the main science data-streams (Table 1).
Table 1: Predicted and measured rates from OBCU assuming 38% effective OBCU dead time (all count/s; for source activity at mission start).
The predictions and observed rates in the two calibration datastreams
agree reasonably well with the predicted rates. These demonstrate that
the calibration lines are being divided almost equally between the
science and calibration data-streams. However, the actual OBCU rate in
the science data represents only a
1% increase over background.
The OBCU events leaking into the science data have a well-defined
spatial and energy distribution and so may be removed during the
uniformity and background correction phase of analysis (Goldwurm 2003).
The ISGRI calibration events (S2 datastream) are received at a rate of 45 c/s. The ISGRI calibration spectrum (Fig. 4)
is dominated by calibration photon induced fluorescence emission from
the tube and hopper, made from lead and tungsten
respectively. Fluorescence is responsible for the complex of lines
around PHA channel
100. The original 511 keV flux is also
detected, if only weakly, at PHA channel
1000 and can be used
for long-term recalibration of both pixel gain and energy-loss
corrections.
![]() |
Figure 4: Typical S2 (ISGRI calibration) spectrum summed over all pixels. PHA values have been gain-normalised, but charge loss correction has not been applied. |
| Open with DEXTER | |
PICsIT calibration events (S5 datastream) are received at a rate of 75 c/s for the detector plane as a whole and stored onboard as histograms
with integration time of typically 1800 s. A typical
pulse-height spectrum is shown as Fig. 5. Note
that the histogramming uses a variable binning to store detailed data
only around the 511 keV (
)
and 1275 keV (
) peaks.
![]() |
Figure 5: Typical S5 (PICSIT calibration) spectrum, summed over all pixels and integrated for one revolution. |
| Open with DEXTER | |
Analysis of PICsIT calibration data may be accomplished at several levels:
- Monitoring of overall detector plane characteristics using the total
spectrum for all pixels. The data quality is good enough to achieve
this at
1 hour timescales.
- Monitoring of 511 keV line position in individual pixels on timescales of 1 revolution.
- Monitoring of 1275 keV line position in individual pixels on timescales of 5 revolutions, allowing an absolute recalibration of pixels gains.
Analysis was performed on PICsIT calibration histogram data throughout the commissioning phase, and the system was seen to perform excellently, tracking the temperature (and hence gain) changes throughout each orbit in full agreement with the temperature values in housekeeping (Malaguti 2003).
The distribution of counts in coincidence with the OBCU across the PICsIT detector plane has been determined by performing background-subtraction on a PICsIT calibration data (S5) histogram. The background in the calibration histogram derives from random coincidences between the OBCU tagging system and other PICsIT events and so can be obtained by scaling of the PICsIT science data for single-site events (S7.0 datastream).
![]() |
Figure 6: Distribution of detected calibration photons across the PICsIT detector plane. The non-uniform distribution derives from a combination of source flux, detection efficiency and tagging efficiency. Note the presence of a few permanently off pixels. |
| Open with DEXTER | |
The calibration count distribution in PICsIT (Fig. 6) shows a number of features:
(a) A clear set of shadows (at y=36, z=15 and z=49) cast by the ISGRI spider. The counting rate in these shadowed regions is reduced almost by a factor of 2.
(b) A region of enhanced counts at y=32, where the edges of the PICsIT modules subtend an increased geometrical area to the OBCU.
(c) A region of reduced counts (y<10). Indeed, there are hints of an overall increasing count rate towards higher y. This is contrary to expectations due to simple distance and/or solid angle effects.
This latter effect was investigated by use of more detailed monte-carlo simulations with a full implementation of the 22Na decay scheme, and much better statistics revealing spatial variations in the calibration photon count distribution. These were performed using the Integral Mass Model (Ferguson 2003).
In data simulated for only OBCU events, the tagging efficiency can be
derived from the fraction of events correctly passed to the S5 calibration
datastream compared to the total counts. Hence:
![]() |
(1) |
A plot of the tagging efficiency derived in this way
(Fig. 7) shows a clear systematic variation across
the detector plane of
10%, with a sharp decline at low y
coordinates. This can be understood by reference to Fig. 1; the volume of the OBCU tagging detector
diametrically opposite the low-y PICsIT pixels is significantly lower
than that for higher-y pixels.
![]() |
Figure 7: Spatial variation of tagging efficiency derived from GEANT4 simulations. |
| Open with DEXTER | |
The IBIS on-board calibration unit (OBCU) is generally working well,
but is suffering from the effect of high-energy particle interactions
which are causing both a blinding (dead-time) effect and a shift in
the recorded pulse height in the active tagging system. The combined
result of these two effects is a reduction of
38% in the
tagging efficiency. This in turn leads to longer times to acquire
acceptable calibration spectra, and some leakage of calibration events
into the science data streams.
For ISGRI, both the direct tagged flux of 511 keV photons and secondary tagged photons (from fluoresence etc) can be used to calibrate the instrument response.
For PICsIT, the time to carry out a pixel-by-pixel recalibration at the required level of better than 1% accuracy is typically 1 orbit for the 511 keV line determination, and 5 orbits for the 1275 keV line. There are no other strong lines present in the PICsIT background, so the use of the OBCU is the only feasible method for PICsIT gain calibration.
For the veto, the line detections in the lateral modules are too weak due to absorption by intervening materials, while the OBCU lines are visible in the rear veto elements. This is the basis of the automatic veto analysis carried out at ISDC.
Acknowledgements
Thanks to the staff at LABEN Spa. for their help in the ground calibration of the IBIS calibration unit. A. J. Bird is funded by PPARC grant GR/2002/00446.